T297 Groundwater Arsenic: A Global Environmental Health Problem and Sustainable Mitigation I
Wednesday, 8 October 2008
8:00 AM - 12:00 PM
Field Deployable Method for Arsenic Speciation in Water
Paper number: T297-8
Program: Technical Sessions
Time: 9:40 AM – 9:55 AM
Authors
-
Lisveth Flores del PinoLa Molina National Agrarian University and Scientific University of the South, Lima, N/A
-
Ivan HavezovBulgarian Academy of Sciences, Sofia, N/A
-
David LongMichigan State University, East Lansing, MI
Abstract
The approach is based on selective retention of arsenic species on specific ion-exchange chromatography cartridges followed by selective elution and quantification using graphite furnace atomic absorption spectroscopy. Water samples can be delivered to a set of three cartridges using either syringes or peristaltic pumps. Species distribution is stable at this point, and the cartridges can be transported to the laboratory for elution and quantitative analysis. A set of ten replicate spiked samples of each compound, having concentrations between 1 and 60 ìg/L, were analyzed. Arsenic recoveries ranged from 78-112 % and relative standard deviations were generally below 10%. Resolution between species was shown to be outstanding, with the only limitation being that the capacity for As(V) was limited to approximately 50 ìg/L. This could be easily remedied by changes in either cartridge design, or the extraction procedure. Recoveries were similar for two spiked hard groundwater samples indicating that dissolved minerals are not likely to be problematic. These results suggest that this methodology can be use for analysis of the four primary arsenic species of concern in drinking water supplies.