FLUORESCENCE OF COMMON pH BUFFERS: A POTENTIAL SOURCE OF MISINTERPRETATION OF GROUNDWATER DYE TRACING RESULTS
Paper number: 249-18
Program: Technical Sessions
Time: 8:00 AM – 6:00 PM
Authors
-
Katherine W. RushGatton Academy, Bowling Green, KY
Abstract
Dyed Oakton® brand pH buffer solutions were quantitatively evaluated for fluorescence in emission wavelengths of typical dyes used in groundwater tracing investigations. All buffers fluoresced. The blue pH 10 had no fluorescence in a range close to any of the dyes considered. The pH 4 (red) buffer emission spectra had a peak emission wavelength within 6.7 nm of that for sulphorhodamine B, and that for the pH 7 (green) buffer was within 4.0 nm of the peak for fluorescein dyes.
While the resulting concentrations were low enough that accidental contamination are unlikely to result in false positives, they are close enough that some ambiguity could be produced. These buffers thus should be carefully kept away from samples and the instrument, in both the field and lab.