POTENTIAL LIMITS ON THE ACCURACY OF PB ISOTOPIC MEASUREMENTS BY LA-ICPMS DUE TO MATRIX EFFECTS

Paper number: T152-15
Program: Technical Sessions
Time: 9:00 AM – 6:30 PM


Authors


Abstract

We present the development of an analytical method for the in situ measurement of Pb isotope ratios in relatively low-concentration samples (~1-40 ppm Pb) using a laser ablation system (193-nm excimer laser from Photon MachinesTM) with a double-focusing single-collector (SC) ICPMS (Nu AttoM ESTM). Our method builds upon a published technique [1,2] that used different LA-SC-ICPMS instrumentation to demonstrate the benefits of fast-scanning ion-counting measurements combined with flat-top peaks. We have paid special attention to the characterization of instrumental artifacts using solutions of the NIST SRM981 Pb isotopic standard in “wet plasma” mode. This includes validating the correction for the interference of 204Hg on 204Pb, characterizing the effects of tails from thallium (with masses 203Tl and 205Tl) on all of the Pb peaks (for Tl/Pb ratios from ~0 to 10 at a constant Pb signal intensity), evaluating the stability of the instrumental mass bias, and maintaining linearity of the detector response over the full dynamic range. Using a suite of international glass standards and newly developed in-house mineral and glass reference materials, we have identified two non-spectral, mass-dependent matrix effects during laser ablation that may limit the accuracy of the measured Pb isotope ratios: (1) one is related to differences in laser spot size and (2) the other is related to differences in the bulk chemistry of the ablated materials (e.g., NIST610 vs. basaltic glasses). The first matrix effect, notable only at the smallest spot sizes (≤20 μm), can presumably be avoided by maintaining a constant spot size for standards and samples. The second matrix effect can potentially be minimized if a standard that closely matches the bulk chemistry of the sample is analyzed. These matrix effects, and the final precision and accuracy of the method, will be discussed in detail at the meeting. [1] Jochum et al. (2005) IJMS 242, 281-289. [2] Jochum et al. (2006) JAAS 21, 666-675.

Geological Society of America Abstracts with Programs. Vol. 48, No. 7

doi: 10.1130/abs/2016AM-280329

© Copyright 2016 The Geological Society of America (GSA), all rights reserved. Permission is hereby granted to the author(s) of this abstract to reproduce and distribute it freely, for noncommercial purposes. Permission is hereby granted to any individual scientist to download a single copy of this electronic file and reproduce up to 20 paper copies for noncommercial purposes advancing science and education, including classroom use, providing all reproductions include the complete content shown here, including the author information. All other forms of reproduction and/or transmittal are prohibited without written permission from GSA Copyright Permissions.