SIMULTANEOUS PB-ISOTOPE AND TRACE-ELEMENT CHARACTERIZATION OF GALENA BY QUADRUPOLE-BASED LA-ICP-MS

Paper number: T152-14
Program: Technical Sessions
Time: 9:00 AM – 6:30 PM


Authors


Abstract

Galena is a common ore mineral in volcanogenic massive sulfide (VMS) deposits, carbonate-hosted Mississippi Valley type (MVT), and diverse vein deposits. It remains widely used for Pb-isotope measurements to discriminate sources of ore fluids in these deposits. Trace-elements in galena analyzed by LA-ICP-MS have also shed light on processes related to episodic fluid input and redistribution of volatile- (Ag, Cd, In, Sb, Tl) and other low-melting-point chalcophile elements during metamorphism and deformation.

Whereas Pb-isotopes in galena are typically measured by TIMS or (LA)-MC-ICP-MS in order to obtain high-precision measurements on individual sample aliquots, Q-LA-ICP-MS could potentially achieve usable precision (e.g, 0.1%) on Pb-isotopes using pooled weighted averages for larger (e.g., n >20) populations. Ablation conditions for galena were optimized to produce a near-steady-state mass transfer to the Q-ICP-MS plasma: 0.35 J/cm2 fluence, 2.5 Hz pulse rate, 60 µm diameter craters and 60 second ablation. The large Pb ion beams produced under these conditions allow short quadrupole dwell times to capture time-dependent fluctuations while at the same time maintaining good counting statistics. All data are externally standardized to a fragment of Broken Hill galena using Pb-isotope reference values of Stacey et al (1969) and trace-element concentrations measured in-house relative to USGS MASS-1. This approach provides near-perfect matrix-matching. A 40s background is sufficient to return to baselines and the high Pb/Hg (typically >1E6 cps) precludes any interference from 204Hg. Several important trace-elements (Ag, Cd, Sn, Sb, and Tl) can also be simultaneously analyzed.

Propagated errors on individual spots is typically between 0.3% to 0.6%. The Isoplot ‘Weighted Average’ routine (weighted by assigned 2σ error, and reported at 95% confidence) reduces these errors to <0.10% for 207Pb/206Pb, ~0.10% for 206Pb/204Pb, 207Pb/204Pb, and 208Pb/206Pb and 0.12% for the 208Pb/204Pb ratios. Although these errors are still significantly larger than TIMS or MC-ICP-MS they are of sufficient precision for rapid reconnaissance studies. The validity of the method was tested by analyzing galena from a wide variety of well-known VMS and vein deposits in Canada, Germany, Ireland, and the USA.


Geological Society of America Abstracts with Programs. Vol. 48, No. 7

doi: 10.1130/abs/2016AM-286686

© Copyright 2016 The Geological Society of America (GSA), all rights reserved. Permission is hereby granted to the author(s) of this abstract to reproduce and distribute it freely, for noncommercial purposes. Permission is hereby granted to any individual scientist to download a single copy of this electronic file and reproduce up to 20 paper copies for noncommercial purposes advancing science and education, including classroom use, providing all reproductions include the complete content shown here, including the author information. All other forms of reproduction and/or transmittal are prohibited without written permission from GSA Copyright Permissions.