CHROMIUM AND NI ISOTOPE ABUNDANCE RATIO MEASUREMENTS IN LOW RESOLUTION USING THE SAPPHIRE DUAL PATH MC-ICP-MS WITH COLLISION/REACTION CELL

Paper number: T36-9
Program: Technical Sessions
Time: 8:00 AM – 5:30 PM


Authors


Abstract

The isotope abundance ratios of the transition metals Cr and Ni provide insights into a wide range of terrestrial processes, and enable environmental and biogeochemical tracing. Moreover, they hold geochronological importance thanks to the presence of short-lived chronometers 53Mn-53Cr and 60Fe-60Ni, which allow accurate dating of early solar system processes. These isotopic systems pose unique challenges due to their small radiogenic variations and the scarcity of precious samples, necessitating high-precision isotope abundance ratio measurements of small sample aliquots.

Plasma-based Cr and Ni isotope measurements suffer from several isobaric interferences such as 40Ar12C+ on 52Cr+, 40Ar14N+ on 54Cr+, or 40Ar18O+ on 58Ni+. To resolve these interferences, both isotope systems are usually measured in medium resolution mode (M/ΔM > 5-95% peak height > ~ 5000), which comes at the cost of reduced sensitivity. In this study, we will present isotope abundance ratio data for Cr and Ni measured in low resolution on a Nu Sapphire Dual Path MC-ICP-MS with Collision/Reaction Cell to remove isobaric interferences.

First results using different gas mixtures (He-H­2, He-O­2, and He-N­2) within the Collision/Reaction Cell reveal residual plasma-based interferences when H2 and O2 are used as reaction gases and potential hydride formation for H2. In contrast, the use of N2 allows for interferences-free measurements of all Cr and Ni isotopes, as evidenced by mass-dependent isotopic variations for both systems. The achieved sensitivity is > 600 V/ppm for either isotope system under dry plasma conditions.

Preliminary consecutive measurements indicate that the 2σ SD short-term repeatability (N = 60) is < 30 ppm for 53Cr/52Cr50/52 and < 70 ppm for 54Cr/52Cr50/52 for repeated 200 s measurements of a 50 ng/ml Cr standard solution. Further data will be presented on the short-term repeatability of Ni isotope abundance ratio measurements, the robustness of the method for either isotope system against concentration mismatching, and the measurement trueness for the isotope standard reference materials NIST SRM 979 (Cr) and NIST SRM 986 (Ni). Results for selected meteoritic samples of known isotopic compositions will be presented as well.


Geological Society of America Abstracts with Programs. Vol. 55, No. 6

doi: 10.1130/abs/2023AM-393875

© Copyright 2023 The Geological Society of America (GSA), all rights reserved. Permission is hereby granted to the author(s) of this abstract to reproduce and distribute it freely, for noncommercial purposes. Permission is hereby granted to any individual scientist to download a single copy of this electronic file and reproduce up to 20 paper copies for noncommercial purposes advancing science and education, including classroom use, providing all reproductions include the complete content shown here, including the author information. All other forms of reproduction and/or transmittal are prohibited without written permission from GSA Copyright Permissions.