IS YOUR ANALYTICAL RESULT ACCURATE?
Paper number: 38-10
Program: THEME SESSIONS
Time: 4:40 PM – 5:00 PM
Authors
Abstract
The calibration curve from 0.5 to 100 ng/L gave a correlation coefficient of 0.999. The instrument responses to low concentrations of mercury in pure water indicate that the method and the laboratory are doing very well.
The Method of Standard Additions (MSA)
In USEPA SW-846 method 7000A, the method of standard additions is described. This technique is best described as a multiple set of spiking experiments using 10, 20, 30, and 40 ng/L spikes of mercury to the sample. When the MSA data are plotted, a correlation coefficient of 0.809 is obtained. These results strongly indicated that the reported value (210 ng/L) was wrong. The MSA experiment shows there is so much positive interference in the method for this sample that the real mercury concentration may be “ND” (non-detected).
Conclusion
All results from any analytical method can be checked for accuracy using the method of standard additions. The MSA is applicable to organic analyses as well as inorganic analyses by any quantitative method.