167 Geochemistry
Tuesday, 18 October 2005
1:30 PM - 5:30 PM
PPB ANALYSIS OF RIVER WATER BY XRF
Paper number: 167-6
Program: Technical Sessions
Time: 2:45 PM – 3:00 PM
Authors
-
Alfred A. MartinRigaku MSC, The Woodlands, TX
-
Takao MoriyamaRigaku MSC, The Woodlands, TX
-
Yasujiro YamadaRigaku MSC, The Woodlands, TX
Abstract
In this study of river water, a newly developed micro-droplet method using "Ultra Carry" is presented. This method implements a new absorbent substance instead of the older paper type filters. The method shows reduced background scattering and improved LLDs.
Calibration lines for Pb, As, Cr, Se, and Cd were generated with standard solutions of river water. In this study, 50mL sample was concentrated to about 1mL with a rotary evaporator, and the concentrated sample was dropped twice at 250uL on the Ultra Carry. The reproducibility tests (n=5) were performed, and standard deviation of a few dozen ppb was obtained for each element (Cr 17ppb, As 14ppb, Se 27ppb, Cd 29ppb, Pb 24ppb).
The newly-developed micro-droplet method, which is easy, simple, and highly reproducible, allows analysis at ppb levels for trace elements in water with a standard XRF spectrometer.